"MARK-10" (Made in USA) Series 5 advanced digital force gauges are designed for tension and compression force testing in numerous applications across virtually every industry, with capacities from 0.12 lb to 2,000 lbF (0.5 N to 10 kN). The gauges feature an industry-leading sampling rate of 7,000 Hz, producing accurate results even for quick-action tests. Accuracy is ±0.1% of full scale, and resolution is 1/5000.
Categories : Inspection Product ,  Force & Torque Instrument ,  Force gauge , 
Brand : MARK-10
Share
Series 5
Advance Digital Force Gauges
Series 5 advanced digital force gauges are designed for tension and compression force testing in numerous applications across virtually every industry, with capacities from 0.12 lb to 2,000 lbF (0.5 N to 10 kN). The gauges feature an industry-leading sampling rate of 7,000 Hz, producing accurate results even for quick-action tests. Accuracy is ±0.1% of full scale, and resolution is 1/5000. A large, backlit graphics LCD displays large, legible characters, while the simple menu navigation allows for quick access to the gauges’ many features and configurable parameters. Data can be transferred to a PC or other data collector via USB, RS-232, Mitutoyo (Digimatic), or analog outputs.
On-board data memory for up to 1,000 readings is included, as are statistical calculations and output to a PC. Integrated set points with indicators are ideal for pass-fail testing and for triggering external devices such as an alarm, relay, or test stand. The gauges are overload protected to 200% of capacity, and an analog load bar is shown on the display for graphical representation of applied force.
The Series 5’s averaging mode addresses the need to record the average force over time, useful in applications such as peel testing, while external trigger mode makes switch activation testing simple and accurate. An ergonomic, reversible aluminum design allows for hand held use or test stand mounting for more sophisticated testing requirements.
Series 5 force gauges are directly compatible with Mark-10 test stands, grips, and software.
The gauges include MESUR® Lite data acquisition software. MESUR® Lite tabulates continuous or single point data. One-click export to Excel allows for further data manipulation.
Features
Ordering Information
Model | Description |
M5-012 | Force gauge, 0.12 lbF / 2 ozF / 50 gF / 0.5 N / 500 mN |
M5-025 | Force gauge, 0.25 lbF / 4 ozF / 100 gF / 1 N / 1000 mN |
M5-05 | Force gauge, 0.5 lbF / 8 ozF / 250 gF / 2.5 N / 2500 mN |
M5-2 | Force gauge, 2 lbF / 32 ozF / 1 kgF / 1000 gF / 10 N |
M5-5 | Force gauge, 5 lbF / 80 ozF / 2.5 kgF / 2500 gF / 25 N |
M5-10 | Force gauge, 10 lbF / 160 ozF / 5 kgF / 5000 gF / 50 N |
M5-20 | Force gauge, 20 lbF / 320 ozF / 10 kgF / 10000 gF / 100 N |
M5-50 | Force gauge, 50 lbF / 800 ozF / 25 kgF / 25000 gF / 250 N |
M5-100 | Force gauge, 100 lbF / 1600 ozF / 50 kgF / 50000 gF / 500 N |
M5-200 | Force gauge, 200 lbF / 3200 ozF / 100 kgF / 1000 N / 1 kN |
M5-300 | Force gauge, 300 lbF / 4800 ozF / 150 kgF / 1500 N / 1.5 kN |
M5-500 | Force gauge, 500 lbF / 8000 ozF / 250 kgF / 2500 N / 2.5 kN |
M5-750* | Force gauge, 750 lbF / 12000 ozF / 375 kgF / 3750 N / 3.75 kN |
M5-1000* | Force gauge, 1000 lbF / 16000 ozF / 500 kgF / 5000 N / 5 kN |
M5-1500* | Force gauge, 1500 lbF / 24000 ozF / 750 kgF / 7500 N / 7.5 kN |
M5-2000* | Force gauge, 2000 lbF / 32000 ozF / 1000 kgF / 10000 N / 10 kN |
Additional Items to Consider | |
15-1004 | MESUR®gauge software |
09-1163 | Cable, Series 5 to RS-232 |
09-1165 | Spare USB cable |
08-1026 | Spare rechargeable battery |
AC1052 | Mounting plate kit |
AC1031 | Spare AC adapter, 220V EUR |
* Note: At 500 lbF (2,500 N) or more, force gauges are not recommended for applications involving sudden, sharp breaks. Examples include compression of brittle materials, pull testing of welds, etc. Consider an indicator with load cell instead.
All models include 110V AC adapters. Add suffix ‘E’ for euro plug (220V),‘U’ for UK plug (220V), or ‘A’ for Australian plug (220V).
Example: M5-50E, M5-200U, M5-1000A
Specifications
Accuracy: | ±0.1% of full scale |
Sampling rate: | 7,000 Hz |
Power: | AC or rechargeable battery. Multi-step low battery indicator is displayed, gauge shuts off automatically when power is too low. |
Battery Life: | Backlight on: up to 7 hours of continuous use Backlight off: up to 24 hours of continuous use |
Measurement units: | lbF, ozF, gF, kgF, N, kN, mN (depending on model) |
Outputs: | USB / RS-232: Configurable up to 115,200 baud. Automatic output available up to 250 Hz. Includes Gauge Control Language 2 for full computer control. Mitutoyo (Digimatic): Serial BCD suitable for all Mitutoyo SPC-compatible devices. Analog: ±1 VDC, ±0.25% of full scale at capacity. General purpose: Three open drain outputs, one input. Set points: Three open drain lines |
Safe overload: | 200% of full scale (display shows “OVER” at 110% and above) |
Weight: | M5-012 – M5-100: |
Environmental requirements: | 40 – 100°F, max. 96% humidity, non-condensing |
Capacity x Resolution
Capacity x Resolution | |||||||
Model | lbF | ozF | kgF | gF | N | kN | mN |
M5-012 | 0.12 x 0.00002 | 2 x 0.0005 | – | 50 x 0.01 | 0.5 x 0.0001 | – | 500 x 0.1 |
M5-025 | 0.25 x 0.00005 | 4 x 0.001 | – | 100 x 0.02 | 1 x 0.0002 | – | 1000 x 0.2 |
M5-05 | 0.5 x 0.0001 | 8 x 0.002 | – | 250 x 0.05 | 2.5 x 0.0005 | – | 2500 x 0.5 |
M5-2 | 2 x 0.0005 | 32 x 0.01 | 1 x 0.0002 | 1000 x 0.2 | 10 x 0.002 | – | – |
M5-5 | 5 x 0.001 | 80 x 0.02 | 2.5 x 0.0005 | 2500 x 0.5 | 25 x 0.005 | – | – |
M5-10 | 10 x 0.002 | 160 x 0.05 | 5 x 0.001 | 5000 x 1 | 50 x 0.01 | – | – |
M5-20 | 20 x 0.005 | 320 x 0.1 | 10 x 0.002 | 10000 x 2 | 100 x 0.02 | – | – |
M5-50 | 50 x 0.01 | 800 x 0.2 | 25 x 0.005 | 25000 x 5 | 250 x 0.05 | – | – |
M5-100 | 100 x 0.02 | 1600 x 0.5 | 50 x 0.01 | 50000 x 10 | 500 x 0.1 | – | – |
M5-200 | 200 x 0.05 | 3200 x 1 | 100 x 0.02 | – | 1000 x 0.2 | 1 x 0.0002 | – |
M5-300 | 300 x 0.1 | 4800 x 2 | 150 x 0.05 | – | 1500 x 0.5 | 1.5 x 0.0005 | - |
M5-500 | 500 x 0.1 | 8000 x 2 | 250 x 0.05 | – | 2500 x 0.5 | 2.5 x 0.0005 | – |
M5-750* | 750 x 0.5 | 12000 x 10 | 375 x 0.2 | – | 3750 x 2 | 3.75 x 0.002 | – |
M5-1000* | 1000 x 0.5 | 16000 x 5 | 500 x 0.2 | – | 5000 x 2 | 5 x 0.002 | – |
M5-1500* | 1500 x 1 | 24000 x 20 | 750 x 0.5 | – | 7500 x 5 | 7.5 x 0.005 | – |
M5-2000* | 2000 x 1 | 32000 x 20 | 1000 x 0.5 | – | 10000 x 5 | 10 x 0.005 | – |
* Note: At 500 lbF (2,500 N) or more, force gauges are not recommended for applications involving sudden, sharp breaks. Examples include compression of brittle materials, pull testing of welds, etc. Consider an indicator with load cell instead.